Performing A Calibration - Lovibond TB 350 Guide De Démarrage Rapide

Masquer les pouces Voir aussi pour TB 350:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 80
Description
1 Application Bar
2 Calibration Curve
3 Standard Set
4 Settings Button
5 Home Button
6 Next Button

4.4.2 Performing a Calibration

Please prepare the T-CAL
calibration. Make sure that the T-CAL
blemishes.
• T-CAL
standards for TB 350 IR: 20 / 800 / 4000 NTU
®
• T-CAL
standards for TB 350 WL: 20 / 800 / 2000 / 4000 NTU
®
1. Zero Adjustment:
Sets zero standard.
2. Prepare Vial:
Oil sample vial if necessary.
3. Insert Standard:
Insert standard value specified on screen.
User will insert series of various turbidity standards to complete
calibration.
4. Modal:
Represents active process.
5. Log Summary:
Press Save Button to save log.
Press Cancel Button to return to beginning of calibration
process.
EN   TB 350
Function
Breadcrumb to help the user through the process steps.
Specifies calibration type.
Unique identifier of standard set.
Used to change settings.
Returns user to home screen.
Progress to next screen.
standards included in the scope of delivery before
®
vials are free of stains, scratches or
®
4 Operation
EN
17