5. Open the lamp housing using the door lock screw
⑤ and remove the lamp holder. (Fig. 19)
6. Remove the plastic block ⑥ from the lamp holder
(or the exhausted lamp in case of replacement)
by loosening the two locking screws ⑦. (Fig. 20)
7. Insert the mercury bulb ⑧ (respect the polarity of
the bulb), tighten the locking screws and refit the
lamp holder into the lamp housing. (Fig. 21)
Disconnect all electrical cables before installing or replacing the bulb.
The bulb has an anode and a cathode of different sizes. Respect the polarity during
assembly, respecting the bulb dimensions.
Do not touch the bulb of the lamp with bare hands to leave no traces of grease on the bulb.
If this happens, clean the bulb with a soft cloth before turning on the lamp.
The lamp has an average life of about 200-250 hours: a time counter and a voltage indicator
are shown on the lamp power supply. Replace the lamp when the hour count exceeds 250
or if the voltage drops below 4.5A.
During use, the lamp, the lamp housing and the surrounding environment become hot.
Before replacing the lamp, switch off the power supply, disconnect all cables and wait for
the lamp and the lamp housing to cool.
After switching on the lamp, wait at least 10-15 minutes before switching it off.
After switching off the lamp, wait for 5-10 minutes before switching it on again so that the
mercury vapors have time to condense.
The lamp contains ultraviolet radiation that could be harmful to eyes and skin. Always look
at the lamp arc through the provided orange screen.