EMC declaration

This product needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
1) Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in
incorrect operation of the unit.
2) Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
3) Caution: this machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked
use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be
used
4) Warning: The use of ACCESSORIES, transducers and cables other than those specified, with the exception of
transducers and cables sold by the MANUFACTURER of the COMPRESSOR NEBULIZER as replacement parts for
internal components, may result in increased EMISSIONS or decreased IMMUNITY of the ME EQUIPMENT or ME
SYSTEM.
Guidance and manufacture's declaration – electromagnetic emission
The ULTRASONIC AIR SALINIZER is intended for use in the electromagnetic environment specified below. The
customer or the user of the ULTRASONIC AIR SALINIZER should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emission
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Guidance and manufacture's declaration – electromagnetic immunity
The ULTRASONIC AIR SALINIZER is intended for use in the electromagnetic environment specified below. The
customer or the user of ULTRASONIC AIR SALINIZER should assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50Hz/60Hz)
magnetic field IEC
61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Compliance
Group 1
Class B
Class A
Complies
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2kV for power
supply lines
supply lines
± 1 kV line(s) to
±1 kV differential
line(s)
mode
<5% U
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 0.5 cycle
for 0.5 cycle
40% U
40% U
T
(60% dip in U
)
(60% dip in U
T
for 5 cycles
for 5 cycles
70% U
70% U
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
<5% U
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 5 sec
for 5 sec
3A/m
3A/m
Electromagnetic environment – guidance
The ULTRASONIC AIR SALINIZER use RF energy only for its
internal function. Its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
The ULTRASONIC AIR SALINIZER is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floor are covered with synthetic material,
the relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
T
)
commercial or hospital environment. If the
T
user of the ULTRASONIC AIR SALINIZER
requires continued operation during power
mains interruptions, it is recommended that
T
)
the ULTRASONIC AIR SALINIZER be
T
powered from an uninterruptible power supply
or a battery.
T
)
T
T
)
T
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment.
6