Celestron 44200 Manuel De L'utilisateur page 15

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Focusing and Changing Power
To achieve the best focus for your micr
adjust the left diopter ring (1) of the lef
th
e zoom power knob (14) at the 0.7x position on the dial which is the lowest power (10x). It is usually best to begin
viewing a specimen at low power, then zoom to high powers. The maximum power is 67x (dial at 4.5x).
1. Turn the focus knob (6) until the specimen comes into focus. If there is not enough travel in the focuser mechanism
to reach focus, it may be necessary to loosen the pillar stand locking knob (5) and raise or lower the entire binocular
head assembly (15) until the image becomes sharp.
2. While looking through the eyepieces, adjust the interpupillary distance by rotating both eyepieces and the eyepiece
holding up and down until the whole field of view can be observed through both eyes at the same time. The reason
this is necessary is that people have varying distance
3. Close your left eye and focus with the right eyepiece until the image is sharp. Then, close your right eye and turn the
left eyepiece diopter ring (1) until the image is in sharp focus through the left eyepiece. Now the interpupillary
distance is set and you can focus only with the focus knob (6).
4. If higher power is desired, rotate the zoom power knob (14) to the desired power. As you increase power, you will
have to refocus the object.
5.
You can adjust the tension of the focus knobs (6) to suit your n
focus knob tight and rotate the right focus knob clockwise to make the tension stiffer. To make the tension looser,
you would rotate the right f
Rep
lacing Illuminator Bulbs
Top Illuminator Bulb – to replace th
1. Remove the top illuminator from
2. Hold the top piece of the illuminator (the piece with the cable attached to it) with one hand. With the other hand, pull
the body away from the top piece by twi
3. Gently remove the bulb from its socket holder.
4. Replace the bulb with a new one of the same type – halogen 12volt, 15watt.
5.
Push the illuminator body up onto the top piece and twist slightly and it will lock in place.
6. Install the illuminator into the bracket arm by screwing
insert.
B
ott m Illuminator Bulb – to replace the bulb:
o
1.
Remove
the stage plate (10) so that you can easily see the fluorescent bulb inside the base (7) – see Figure 4d.
2. Raise the head (15) up so you have more room to work in.
3. Gently remove the illuminator bulb from the bu
4. Replace the bulb with a new one of the same type – fluorescent 8watt and either 115-volt or 230-volt depend
the model you have. Put the bulb in gently so the pins of th
Figure 4c
R
eplacing Fuses
Two
fuses are located on the back of the microscope (see Figure 4b). They are T1A fuses. If the power goes out or will not
come on, it is possible
one or both of the fuses needs to be replaced. Make sure the power switch is turned off and then use
a flat head screwdriver to rem
b
ox back in place.
oscope, not only can you make adjustments with the focus knobs, but you can also
t eyepiece as well as the interpupillary distance between the eyepieces: Start with
ocus knob counterclockwise.
e bulb:
the bracket arm by unscrewing the illuminator lock screw (4).
sting slightly and the bulb can be seen in its socket holder – see Figure 4c.
lb holder by pulling outward and moving it left to right gently.
ove the fuses. If they are blown out, replace them with new fuses and then install the fuse
s between their eyes.
eeds. Looking straight at the microscope, hold the left
the illuminator lock screw (4) into the illuminator threaded
e bulb insert properly into the bulb socket holder.
Figure 4d
15
ing on

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

442044420244206

Table des Matières