Model C, model H (seperated sensor probe)
For mounting template for metal and polycarbonate housing see page 17.
Transmitter EE23-xCx and EE23-xHx series are with seperated sensor probe.
working range sensor probe:
working range electronics:
Mounting of the sensor probe - model C (12mm):
Using the stainless steel mounting flange (refer to accessories) it is pos-
sible to mount the probe on the outer wall of the measuring chamber.
The depth of immersion is adjustable.
For roof installations use the drip water protection (refer to accessories) to
protect the sensor head and elements against condensed water.
The transmitter probe must point horizontal or downwards. The
cable should hang loosely (bend for draining water of condensation).
Mounting of the sensing probe - model H (5mm):
Incl. mounting fixture for concealed mounting or in tight spaces
(e.g. in museums, door frame, etc...).
bore diameter: 13mm
(0.5")
material thickness: min. 3mm
Model F (wall mounting type with rear cable outlet)
For mounting template for metal and polycarbonate housing see page 17.
Positioning of bore holes top left to center line of the probe:
metal housing:
polycarbonate housing: x = 20.5mm
Transmitter EE23-xFx series are designed for wall mounting in clean room
applications. Connection cables do not disturb the cleaning process.
- working range sensor probe:
- working range electronics:
The transmitter must be mounted with the sensor probe pointed
downwards.
Model G (wall mounting - for meteorology)
For mounting template for metal and polycarbonate housing see page 17.
Transmitter EE23-xGx series are designed for outdoor applications.
- working range sensor probe:
- working range electronics:
For outdoor applications the transmitters must be equipped with a radiation
shield (refer to accessories). The shield causes forced ventilation which
prevents overheating of the sensor probe in direct sunlight. It also protects
the sensing probe against water of direct rainfall.
The sensor probe must point upwards and should be used in
combination with the radiation shield (refer to accessories).
18
model C: -40...120°C
model H: -40...80°C
-40...60°C
(0.1")
x = 28.5mm
(1.1")
(0.8")
-40...60°C (-40...140°F)
-40...60°C (-40...140°F)
-40...60°C (-40...140°F)
-40...60°C (-40...140°F)
(-40...248°F)
(-40...176°F)
(-40...140°F)
y = 37.5mm
(1.5")
y = 25.4mm
(1