- Only type-tested batteries may be used.
- Avoid use in aggressive acids or bases.
- It must be ensured that the unit is not taken into zone 0.
5. Ex-Data
EC type approval test certificate no.: PTB 01 ATEX 2018
Ex-designation:
Approved for zone 1, device group II, explosion group IIC
(explosive gases, vapours or mist), temperature class T4.
Report Job Identification Nr.: 3009962
FM-Indicator: Class I Zone 0 AEx ia IIC T6 Ta = 50 °C
Permitted for Zone 0, Equipment group ii, Gas group C,
Temperature Class T6
Intrinsically safe for use in a Class I, Division 1, Groups A, B,
C & D with a Temperature Class of T6
6. Technical specifications
Detection:
Detectable
magnetic:
Indicator:
Ambient
temperature Ta:
Storage
temperature:
Batteries /
power supply:
(Table with type-tested batteries)
1
2
3
4
5
6
7
II 2 G Ex ia IIC T4
I.S. Class I Division 1 Group A-D T6 Ta = 50°C
no contact with test object required
alternating, direct and permanent fields
optical, via built-in LED
-20 °C ... + 50 °C
-40 °C ... + 60 °C (without batteries)
2 x LR03 according to IEC, AAA
Manufacturer:
Varta
Varta
Eveready
Duracell
Duracell
Varta
Energizer
Type:
LR03 High Energy No.4903
LR03 Industrial Alkaline No. 4003
Energizer No. E92
Plus MN2400 LR03
Procell MN2400 LR03
LR03 Maxi Tech No.4703
Ucar Gold LR03
13