Guidance and manufacturer's declaration – electromagnetic emissions
The ME equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME equipment should ensure that it is used in such an environment.
Emissions Test
Compliance
RF Emissions
The ME equipment uses RF energy only for its internal function. Therefore, its RF emissions are
Group 1
CISPR 11
very low and are not likely to cause any interference in nearby electronic equipment.
RF Emissions
Class B
CISPR 11
Harmonic emissions IEC
Not Applicable
61000-3-2
Voltage fluctuations /
Not Applicable
flicker emissions
Non-Life Support Equipment Separation Distance Calculation (3Vrms / 3V/m compliance)
Separation distance according to frequency of transmitter (m)
Rated maximum output power of
150 kHz to 80 MHz in ISM bands
transmitter (W)
3,5
d = [
] P
V
1
0.01
0.12
0.1
0.37
1
1.17
10
3.69
100
11.67
Guidance and manufacturer's declaration – electromagnetic immunity
The ME equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME equipment should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge (ESD)
±6kV Contact
IEC 61000-4-2
±8kV Air
Radiated RF
3 V/m 80MHz to 2.5GHz
IEC 61000-4-3
Conducted RF
3Vrms 150kHz to 80MHz
IEC 61000-4-6
Electrical fast transient
±2kV power line
IEC 61000-4-4
±1kV I/O lines
Surge
±1kV differential
IEC 61000-4-5
±2kV common
Power frequency magnetic field
3 A/m
IEC 61000-4-8
>95% dip 0.5 cycle
Voltage dips, short interruptions
and voltage variations on power
60% dip 5 cycles
supply input lines
70% dip 25 cycles
IEC 61000-4-11
95% dip 5 sec.
Electromagnetic environment – guidance
Complies
The ME equipment is
solely battery powered.
80 MHz to 800 MHz
800 MHz to 2.5 GHz
3,5
7
d = [
] P
d = [
] P
E
E
1
1
0.12
0.23
0.37
0.74
1.17
2.33
3.69
7.38
11.67
23.33
Compliance level
Electromagnetic environment – guidance
Floors should be wood, concrete, or ceramic tile. If
Complies
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Field strengths outside the shielded location from
Complies
fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than 3
V/m.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Not Applicable
(no electrical cabling)
Separation distance calculation provided above. If a
known transmitter is present the specific distance
can be calculated using the equations.
Not Applicable
The ME equipment
is solely battery powered.
Not Applicable
Power frequency magnetic fields should be at levels
Complies
characteristic of a typical location in a typical
commercial or hospital environment.
The ME equipment
Not Applicable
is solely battery
powered.