Using The Jb Aqua Plus; Bath Controls; Indicator Lamps; Setting The Control Temperature) - Grant JB Aqua Plus Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

5.2

Using the JB Aqua Plus

5.2.1 Bath controls

Sample
protection
thermostat
Warning light
Power on light

5.2.2 Indicator lamps

There are three indicator lights:
1. Power on (green).
2. Heater on (orange). Marked
3. Warning (orange). Marked indicates that the temperature is being controlled by the
sample protection thermostat.
5.2.3 Setting the control temperature ()
The water temperature of the bath can be set using the larger temperature control knob.
1. Turn the knob of the sample protection thermostat fully to maximum.
2. Turn the knob of the temperature control to the desired temperature.
3. Turn on the unit using the switch on the rear of the bath and wait until the
temperature has stabilised.
4. Measure the temperature with a thermometer and adjust the temperature control
knob if necessary to obtain the desired temperature.
5. Set the sample protection thermostat as detailed below if needed.
29825 Ver. 2 - July 2010 71W
Page 7
Heater on light
Temperature control
.
www.grant.co.uk
SUB, JB & SBB Aqua Plus
Operating Manual

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sub aqua plusSbb aqua plus

Table des Matières