After filling the bath adjust the water flow rate to the minimum which maintains a constant
water level when the water is boiling. Check that any relevant local water supply
regulations are complied with if connecting to a mains water supply.

5.4.5 Resetting the over temperature cut-out devices

Two fixed over temperature cut-outs prevent the heater from overheating in the case of a
low water level. The reset buttons for the over temperature cut-outs are on the side panel
and can be reset by removing the protective black cap and pressing the red button in.
The over temperature cut-outs may activate during normal operation. If this is the case
then they can be reset and the bath can continue to be used without compromising the
safety of persons or the surroundings. However, if the cut-outs are persistently activated
then it is recommended to have the bath checked by a competent person as soon as
possible.
6.0

Technical specifications

Operating conditions
Ambient Temperature
Maximum relative humidity
Altitude above sea level
Operating Enviroment
Electrical details
Mains supply:
230V @ 50/60 Hz or 120V @ 50/60 Hz
Polution degree:
2
Installation Category: II
Note: Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply
voltage
Models
SUB Aqua 2 Plus
SUB Aqua 2S Plus
SUB Aqua 5 Plus
SUB Aqua 12 Plus
SUB Aqua 18 Plus
SUB Aqua 26 Plus
SUB Aqua Dual Plus
SUB Aqua 34 Plus
JB Aqua 2 Plus
JB Aqua 2S Plus
JB Aqua 5 Plus
JB Aqua 12 Plus
JB Aqua 18 Plus
JB Aqua 26 Plus
JB Aqua Dual Plus
SUB, JB & SBB Aqua Plus
Operating Manual
5 to 40°C
80% R.H. in room temperatures up to 31°C
decreasing linearly to 50 % R.H. at 40°C
Up to 2,000 m (6,500 ft)
Indoor use only
Capacity (L)
2
2
5
12
18
26
5 & 12
34
2
2
5
12
18
26
5 & 12
www.grant.co.uk
Current Rating (A)
120V
230V
1.0
0.5
2.6
1.3
2.6
1.3
5.3
2.7
7.9
5.3
7.9
5.3
7.9
4.0
10.5
8.5
1.0
0.5
2.8
1.3
2.8
1.3
5.6
2.7
8.4
5.3
8.4
5.3
8.2
4.0