EMC data
Guidance and declaration - electromagnetic immunity - The model MyOxy is intended for
use in.the electromagnetic environment specified below. The customer or the user of the model
MyOxy should assure that it is used in such an environment.
EN
60601
Immunity test
test
level
3 Vrms
150 kHz
Conducted RF
to 80
IEC
MHz
61000-4-6
Radiated RF
3 V/m
IEC
80 MHz
61000-4-3
to 2.5
GHz
NOTE I At 80 MHz end 800 MHz. the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
40
Compliance
level
Portable and mobile RF communications
equipment should be used no closer to any part
of the model MyOxy, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d=1.2×P½
d=1.2×P½ 80 MHz to 800 MHz
d=2.3×P½ 800 MHz to 2.5 GHz
3 V
where P is the maximum output power rating
of the transmitter In watts (W) according to
3 V/m
the transmitter manufacturer and d Is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a
should be less than the compliance level in each
frequency range.b
Interference may occur In the vicinity of
equipment marked with the following symbol:
BewellConnect - BW-OX1 - User Manual - 072017v2
Electromagnetic
environment - guidance
F
d
t
t
N
N
a