shorted condition. The resistors are
rated to 140 +/- 10%. Fusing current
is 1A at 125V. Fusing time is 30-60ms
at 125V, without causing dangerous
conditions during opening of the fuse
element. Leads need to be replaced
when fuse has opened.
Note that TL-80 and TL-82 test leads
can only be specifically used with the
61-085 and 61-086 testers.
Maintenance:
• Clean the case with a damp cloth and
mild detergent. Do not use abrasives or
solvents.
Service, and Replacement Parts:
For replacement parts or to inquire about
service information contact
IDEAL INDUSTRIES, INC.
Attn: Repair Dept.
1000 Park Avenue
Sycamore, IL 60178
Technical Support: 1-877-901-2005
or visit our website
www.testersandmeters.com.
Specifications:
VAC Ranges: 120V, 240V, 480V, 600V AC.
VAC Accuracy: Relative indication only
VDC Ranges: 120V, 240V, 600V DC
VDC Accuracy: Relative indication only
Continuity (61-086 only): Continuous LED
and audible indication at less than 500kΩ
- 11 -
resistance.
Response time of 100ms.
Operating Frequency: 25-60 Hz
Input Impedance: Low
Operating Environment: 32° F to 122° F
(<70% humidity)
Storage Temp.: -4° F to 140° F
(<80% humidity)
Altitude: 2000m
Indoor Use
Battery: (4) 1.5V (IDEAL# 61-201, IEC
#LR44, or NEOA# 1166A).
Battery Life: 200 hours typical.
Accessories included: Test Leads, (4)
1.5V "LR44" batteries, operating instructions
Dimensions: 6.0" (H), 2.3" (W), 1.2" (D)
Weight: 8.0 oz
Safety: EN61010, CAT III 600V
Double Insulation
Warranty Statement:
This tester is warranted to the original
purchaser against defects in material and
workmanship for two years from the date of
purchase. During this warranty period,
IDEAL INDUSTRIES, INC. will, at its option,
replace or repair the defective unit, subject
to verification of the defect or malfunction.
- 12 -