Technical data

(Technical revisions reserved. 21W12)
Infrared temperature
Display resolution
Ambient temperature /
wet-bulb temperature
Relative humidity
Dew point temperature
Contact temperature,
K-type
Optics
Emission coefficient
Laser
Laser wavelength
Laser class
Power supply
Runtime
Operating conditions
Storage conditions
Radio module
operating data
Dimensions (W x H x D)
Weight (incl. batteries)
EU directives and disposal
This device complies with all necessary standards for the free movement
of goods within the EU.
This product is an electric device and must be collected separately for disposal
according to the European Directive on waste electrical and electronic equipment.
Further safety and supplementary notices at:
http://laserliner.com/info?an=cosppl
-40°C ... 800°C
-40°C ... 0°C (± (1°C + 0.1°C / 1°C))
0°C ... 33°C (± 1°C or ± 1%
depending on the higher value)
>33°C (± 2°C or ± 2%
depending on the higher value)
0.1°C / 0.1% rH
-20°C ... 65°C
0°C ... 50°C (± 1°C)
<0°C and >50°C (± 2.5°C)
1% ... 99%
20% ... 80% (± 3%)
<20% and >80% (± 5%)
-50°C ... 50°C
41% rH ... 95% rH (± 1.5°C)
31% rH ... 40% rH (± 2°C)
20% rH ... 30% rH (±2.5°C)
-30°C ... 1372°C (± 1°C or ± 1%
depending on the higher value)
30:1 (30 m distance : 1 m measured spot)
0.01 - 1.0 adjustable
8-point laser circle
650 nm
2 < 1 mW
Batteries, 2x 1.5 V type AA
20 hours
0 ... 50°C, 80% rH, no condensation,
max. altitude 2000 m
-10 ... 60°C, 80% rH, no condensation
IEEE 802.15.4. LE ≥ 4.x (Digital Connection) interface; Frequency band: ISM band
2400–2483.5 MHz, 40 channels; Transmission power: max. 10 mW; Bandwidth: 2 MHz;
Bit rate: 1 Mbit/s; Modulation: GFSK/FHSS
150 x 205 x 60 mm
412 g
CondenseSpot XP
-40°F ... 1472°F
-40°F ... 32°F (± (1.8°F + 0.18°F / 1°F))
32°F ... 91,4°F (± 1.8°F or ± 1%
depending on the higher value)
>91.4°F (± 3.6°F or ± 2%
depending on the higher value)
0.1°F / 0.1% rH
-4°F ... 149°F
32°F ... 122°F (± 1.8°F)
<32°F and >122°F (± 4.5°F)
-58°F ... 122°F
41% rH ... 95% rH (± 2.7°F)
31% rH ... 40% rH (± 3.6°F)
20% rH ... 30% rH (± 4.5°F)
-22°F ... 2501.6°F (± 1,8°F or ± 1%
depending on the higher value)
32 ... 122°F, 80% rH, no condensation,
max. altitude 2000 m
14 ... 140°F, 80% rH, no condensation
25