Materials in Direct Contact with the
Operator
• Concentrator Casing ......................................................................................................................PC/ABS.
• Concentrator Control Panel ..........................................................................Polyester EBG7 or equivalent
• Concentrator Handle .............................................................................................................Polycarbonate
• Oxygen Outlet, Nozzle ..........................................................................................................Polycarbonate
• Unit Label ..........................................................................................................................................Lexan
• Cord Connectors ...........................................................................................Polycarbonate/Vinyl chloride
• Power Cord .............................................................................................................................. PVC, Metal
• Power Supply .................................................................................................. .Lexan 940 (Polycarbonate)
• Battery Pack ...................................................................................................................................PC/ABS
• Battery Pack, Power Supply Labels ...................................................................................... Polyester film
• Concentrator Carrying Case .................................................................. Cordura, Neoprene, Nylon Lining
• Strap ............................................................................. Polypropylene, Acetyl, Vinyl, Cordura, Neoprene

EMC Testing

Guidance and Manufacturer's Declaration —Electromagnetic Emissions
The FreeStyle and FreeStyle 5 are intended for use in the electromagnetic environment specified below. The
customer or the user of the FreeStyle and FreeStyle should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Electromagnetic Environment - Guidance
The FreeStyle and FreeStyle 5 uses RF energy only for its internal
Group 1
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
Class B
The FreeStyle and FreeStyle 5 are suitable for use in all
establishments, including domestic establishments and those
Class A
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Complies
FreeStyle & FreeStyle 5 User Manual — 27