regulation can be provided by an external
regulator.
In the case of the E series, temperature
setting is regulated by the analogue dial and
the required level is determined by using a
thermometer
(not
supplied).
This
also
applies to the EAM series with individual
settings for each position.
For the CM series the temperature is
reached by setting the dial to the required
level and checked with a thermometer if
required.
Monitoring
and
stability
are
achieved with the K type probe which is
plugged into the rear of the unit.
For the DM series the temp level is set by
use
of
the
digital
potentiometer
and
monitored by the K type probe.
The ES series Stirrer Mantles have an
additional speed (RPM) setting.
During the heating stage on the CM, E,
EAM, ES series the yellow "Heating" light will
come on periodically. The frequency of the
heating
mode
will
depend
on
the
temperature setting.
Turn off the heating controllers before
removing the flasks and when the process is