Télécharger Imprimer la page

Nahita 5134IN Serie Mode D'emploi page 25

Publicité

Les langues disponibles

Les langues disponibles

Weighing unit selection
Press PRT key to select the desired weighing unit among mg (UNT-0), g (UNT-1), ct
(UNT-2) and oz (UNT-3).
The weighing unit by default is g.
Piece counting function
1. Keep pressed ON key until Set-0 is displayed.
2. Then, press PRT key to select the reference number of pieces (COU-10, COU-25
or COU-50).
3. Put the corresponding number of pieces on the pan, press CAL key and wait until
the number of pieces is displayed.
4. Remove the pieces from the pan and the balance is ready to be used in piece
counting mode (weighing unit will change from g to pcs).
Note: In counting mode, the weight of the pieces must be even and the weight of an
individual piece should not be less than the readability of the balance.
To return to weighing mode:
1.
Keep pressed ON key until Set-0 is displayed.
2.
Press PRT key to select COU-00
3.
Press TAR key and the balance will enter weighing mode.
Percentage function
1.
Keep pressed ON key until Set-0 is displayed.
2.
Then, press PRT key to select the parameter COU-100.0%.
3.
Put on the pan the reference simple, press CAL and wait until 100.0% is
displayed.
4.
Remove the sample from the pan and the balance is ready to be used in
percentage mode (weighing unit will change from g to %).
To return to weighing mode:
1.
Keep pressed ON key until Set-0 is displayed.
2.
Press PRT key to select COU-00
3.
Press TAR key and the balance will enter weighing mode.
25

Publicité

loading

Ce manuel est également adapté pour:

5134ex serie5133