Télécharger Imprimer la page

Nahita 5134IN Serie Mode D'emploi page 24

Publicité

Les langues disponibles

Les langues disponibles

display; after releasing the keys CAL-200 (value of the necessary weight to make
external calibration) will be displayed.
2.
Put the calibration weight on the pan and wait until its weight is displayed (e.g.
200.0000g).
3.
Remove the weight and the balance will enter weighing mode.
Note: Each time the external calibration is made it is necessary to make internal
calibration first.
External calibration models (series 5134EX and 5133)
1.
Press CAL key; the value of the necessary weight will flash on display.
2.
Put the calibration weight on the pan; for this, open the glass door, putt he weight
on the pan and close the door again. Wait until the value of calibration weight
stops flashing and is fixed on display (e.g. 200.0000g).
3.
Remove the weight from the pan; for this, open the glass door, remove the weight
from the pan and close the door again.
4.
The balance will enter weighing mode and will display 0.000g or 0.0000g.
Note: Each balance includes the necessary weight to make external calibration.
Tare function
1.
Put the recipient on the pan; the corresponding weight will be displayed.
2.
Press TAR key and Reading on display will automatically set to 0.000 g or 0.0000
g; tare is complete.
4.
If the recipient is removed from the pan, the corresponding weight with negative
value will be displayed.
6. Press TAR key again and the value on display will be set to zero.
24

Publicité

loading

Ce manuel est également adapté pour:

5134ex serie5133