Télécharger Imprimer la page

Use Of The Balance - Nahita 5134IN Serie Mode D'emploi

Publicité

Les langues disponibles

Les langues disponibles

nothing has been damaged:
Balance
-
Pan
-
Power cable (AC adapter)
-
Instruction manual
-
External calibration weight
-
In case the balance or any component presents any damage due to transport tell it
immediately to your transport agent or dealer so that they can make the claims in the
correct time limit.

4. USE OF THE BALANCE

Level adjustment
Once the balance is placed on its location, check the bubble level; if bubble is not well
centered, turn the adjustable feet so as the bubble moves to the center and the
balance is perfectly leveled.
Start
1.
Connect the balance to the power supply and press ON key on the panel.
2.
When the balance turns on, a countdown of 30 min will be displayed showing the
necessary pre-warming time for a better performance and stable state of the
balance.
3.
Once pre-warming time has passed it is recommended to calibrate the balance.
In internal calibration models once pre-warming time has passed, the balance will
automatically make calibration.
4.
After calibration, the balance will enter weighing mode.
5.
To turn the balance off, press OFF key on the panel and display will light off. If the
balance will not be used for a long time, disconnect it from the power supply.
Calibration
Internal calibration models (series 5134IN)
1. Press CAL key
2. Calibration using the internal weight will be made automatically.
If tolerance (error) after internal calibration is too high, then it is necessary to make
external calibration. For this, proceed as follows:
1.
Press simultaneously CAL and ON keys until F" appears at the upper side of
23

Publicité

loading

Ce manuel est également adapté pour:

5134ex serie5133