Thermometer Operating Instructions
The thermometer is a non-contact infrared thermometer. There are many mathematical modes for the Infrared function.
Please remember to keep away from children and don‟t use it for safety related applications.
(Default Screen)
Simply aim the thermometer at the measure target
with Lens (9) and press Meas. key (7) to display
the surface temperature. The Distance:Spot is
12:1. Please make sure the target area is within the
field of view.
FUNCTION
Press Mode key (2) for scrolling more display function as follows.
Here will show the emissivity data. (The default emissivity is 0.95.)
Press Mode key (2) , then press Lock key (4) or
confirm it. The emissivity can be changed from 0.10 (10E) to 1 (100E)
Press Mode key (2), for Maximum (MAX), Minimum (MIN), die Difference between MAX and
MIN (DIF) and Average (AVG) modes. During measurement, the special modes reading will be
displayed beside the mode icon.
o
o
Press Lock key (4) oder
C/
then press Meas. key (7) to confirm it. When the reading is outside the High Alarm (HAL) or Lo
Alarm (LAL) limit, the High or Low icon will flash and you will hear a beep sound.
Xylem Analytics Germany GmbH
Am Achalaich 11
82362 Weilheim
Tel.: +49 841 95478-0
Fax: +49 841 95478-80
E-Mail: ebro@xylem.com
Internet: www.ebro.com
* IP54 Splash Proof
o
o
C/
F key (3) to set the emissivity, then press Mode key (2) to
F key (3) to change the High Alarm (HAL) or Lo Alarm (LAL) ,
Add Value
In E,MAX, MIN, DIF, AVG
mode:
In all modes: First hold on the
Meas. key (○,7)
Memory
CAUTION
1.
WHEN THE DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAM - PERMANENT EYE DAMAGE
MAY RESULT.
2.
USE EXTREME CAUTION WHEN OPERATING THE LASER.
3.
NEVER POINT THE DEVICE TOWARDS ANYONE'S EYES.
4.
KEEP OUT OF REACH OF ALL CHILDREN.
STORAGE & CLEANING
It should be stored at room temperature. The sensor lens is the most delicate part of the thermometer. The lens should be kept
clean at all times, care should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical
alcohol, allowing the lens to fully dry before using the thermometer. Do not submerge any part of the thermometer.

LCD ERROR MESSAGES

The thermometer incorporates visual diagnostic messages as follows:
'Hi' or'Lo' is displayed when the temperature being measured is outside the settings of HAL and LAL.
'Er2' is displayed when the thermometer is exposed to rapid changes in the ambient temperature. 'Er3' is displayed
when the ambient temperature exceeds 0°C (32°F) or +50°C (122°F). The thermometer should be allowed plenty
of time (minimum 30 minutes) to stabilize to the working/room temperature.
Error 5~9, for all other error messages it is necessary to reset the thermometer. To reset it, turn the instrumet off,
remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If the error message
remains please contact the Service Department for further assistance.
Batteries
The thermometer incorporates visual low battery indication as follows:
Battery OK: measurements are
possible
Battery replacement
1. Pull out the rubber gasket (10)
2. Loosen the screw and pull out the battery cover
3. Replace and reinstall with new batteries
4. Place the battery cover back and fasten the screw
5. Plug the rubber gasket (10) to avoid failure of water proof
When the „Low Battery‟ icon indicates the battery is low, the battery should be replaced immediately with AAA,
1.5V batteries. Please note: It is important to turn the instrument off before replacing the battery otherwise the
thermometer may malfunction.
Dispose of used battery promptly and keep away from children.
03/2022
ba80038def02_TFI-54
Press Lock key (4) for LOCK mode ON/OFF. The lock mode is particularly useful for
continuous monitoring of temperatures for up to 60 minutes.
o
o
o
o
Press
C/
F key (3) for
C or
F transferred.
and press Lock key (4) for backlight function ON/OFF.
and press oC/ oF key (3) for laser function ON/OFF.
Recall the last memory by pressing Mode key (2) while power off
Battery Low: battery needs to be replaced,
measurements are still possible
Battery Exhausted: measure-
ments are not possible