ELECTRICAL AND ELECTROMAGNETIC
SAFETY
Guidance and manufacturer's declaration - electromagnetic emissions
Ted K-Ortho is intended for use in the electromagnetic environment specified
below. The customer or the user of Ted Orthopedics should assure that it is used
in such an environment.
Emissions test
Conformity
Emissions
Groupe 1
RF* CISPR 11
Emissions
Class B
RF CISPR 11
Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
Ted Orthopedics is intended for use in the electromagnetic environment
specified below. The customer or the user of Ted Orthopedics should assure
that it is used in such an environment.
RF* : Radio-frequency
36
Electromagnetic environment -
guidelines
Ted Orthopedics uses RF energy only for
its internal functions. Therefore, its RF
emissions are very low and are unlikely
to cause interference in nearby electronic
devices.
Ted Orthopedics is intended for office or
home use.
Test
Test level
immunity
CEI 60601
Landfills
±
8
kV
in
electrostatic
contact
conformity
(DES)
± 15 kV in the
CEI 61000-4-2
air
Disturbances
10V/m de 80
Radiated RF
MHz à 2,7 GHz
CEI 61000-4-3
Disturbances
380 - 390 MHz
Radiated RF
27 V/m ; PM
CEI 61000-4-3
50% ; 18 Hz
Fields
430 - 470 MHz
near RF wireless
28 V/m (FM
communication
±5 KHz, 1 Khz
equipment
sine)
PM ; 18 Hz
704 - 787 MHz
9V/m; PM 50%
; 217 Hz
Electromagnetic environment
Level
- guidelines
Floors should be wood, concrete
or ceramic tiles. If floors are
covered with synthetic materials,
the relative humidity should be at
least 30%.
10 V/m
-
Communication devices should
not be used (including peripherals
27 V/m
such as antenna cables and external
antennas)
28 V/m
9 V/m
37