4. Specifications

The ND 140x QUADRA-CHEK is an advanced evaluation unit for performing 2, 3 and 4
axis measurements at very high levels of precision and accuracy. This unit is designed for
indoor use only. The ND 140x components shall only be installed as described in these
instructions. Mounting, installation and maintenance are to be performed by qualified
personnel only.
Power input:
Replaceable fuse:
Environmental conditions
Operating temperature:
Storage temperature:
Relative humidity:
Altitude:
Protection:
Overvoltage category:
Pollution degree:
Mass:

5. Mounting

The following items are included with the product:
• Power cord
• Stand or mounting adapter
• 2-D demo part
• ND 1400 QUADRA-CHEK Installation Instructions
The following accessories are available for this product:
• Foot switch, ID 681041-01
• Stand, ID 382892-02
• Mounting adapter, ID 682419-01
• 2-D demo part, ID 681047-02
• 3-D demo part, ID 681048-01
Refer to " Accessories and connector pin assignments" on page 116 for additional
information. Refer to the instructions provided with an accessory for additional
information.
Do not mount in a position that makes it difficult to access the power switch or
power cable.
The ND 140x can be used as a tabletop unit when placed on the provided stand. The
unit can then be tilted forward or backward by up to 20° for the best reading angle. The
ND 140x is secured to the swivel slots of the stand by a shoulder screw (5), a cap screw
(1) and washers (2,
3
& 4).
To attach the stand:
Attach the unit to the stand as shown at the right
X
X
Partially tighten the cap and shoulder screws
X
X
Adjust the unit to the desired tilt position
X
X
Completely tighten the cap and shoulder screws
X
X
Benchtop location and mounting
Rest the unit on a flat, stable surface or bolt the unit to a stable surface from the bottom
using four M5 screws fastened in the pattern shown in "Dimensions" on page 115.
AC 100 V ... 240 V (-15 % ... +10 %)
50 Hz ... 60 Hz (±3 Hz)
(max. 100 W)
T 1.6 A / 250 V, 5 mm x 20 mm
Quantity: 2
0 °C ... 45 °C
-20 °C ... 70 °C
† 80 %
† 2000 m
IP00, front IP40
II, equipment intended to be supplied from the building
wiring
2
ND with stand: approx. 4.8 kg
ND with mounting adapter: approx. 2 kg
4
1
2 3
Stand assembly