Output Data - Knick ThermoTrans A20210P0/ Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Specifi cations

Output Data

Outputs
Control range
Resolution
Load
Current output
Voltage output
Output error limits
Current output
Voltage output
Residual ripple
Current output
Voltage output
Temperature coeffi cient at
the output
Error signaling
46
0 ... 20 mA, 4 ... 20 mA,
0 ... 10 V, or 0 ... 5 V,
calibrated switching
0 % to approx. 102.5 % span
at 0 ... 20 mA, 0 ... 10 V or
0 ... 5 V output
-1.25 % to approx. 102.5 % span
at 4 ... 20 mA output
16 bits
≤ 500 Ω
≥ 10 kΩ
± (10 μA + 0.05 % meas.val.)
± (5 mV + 0.05 % meas.val.)
< 10 mVrms (at 500 Ω load)
< 10 mVrms (at 10 kΩ load)
50 ppm/K of end value (average TC
in permitted operating temp range,
reference temp 23 °C)
Output: 4 ... 20 mA:
Current ≤ 3.6 mA or ≥ 21 mA
(see table on page 42 for more data)

Publicité

Table des Matières
loading

Table des Matières