Battery and Fuse Replacement

Refer to the following figure to replace the fuse and batteries :
CAUTION
• Use only a fuse with the amperage, interrupt, voltage and speed
rating specified.
• Fuse rating:
440mA,1000V IR 10KA Fuse (size 35 x 10mm)
11A 1000V IR 20KA Fuse (size 38 x 10mm)
• To avoid a false reading, replace the batteries as soon as the low
battery indicator
appears.
• 1.5V x 4 alkaline batteries.
Maintenance
Clean the case with a damp cloth and mild detergent. Do not use
abrasives or solvents.
Service and Replacement Parts
No user-serviceable parts.
For replacement parts or to inquire about service information, contact
IDEAL INDUSTRIES, INC. at 1-877-201-9005 or visit our website @
www.testersandmeters.com.
17
Specifications

General Specifications

Maximum voltage applied
1000V ac rms or dc
to any terminal:
Display
4000/40000 counts for 61-497
10000/100000 counts for 61-498
Polarity Indication:
Automatic, positive implied, negative polarity
Batteries Life:
100 hours
Low Batteries Indication:
"
drops below operating voltage
Low battery voltage:
Approx. 4.5V
Auto Power Off:
Default 30 minutes
Operating Ambient:
Non-condensing ≤ 10°C
11°C ~ 30°C (≤ 80% RH),
30°C ~ 40°C (≤ 75% RH),
40°C ~ 50°C (≤ 45% RH),
Storage Temperature:
-20°C to 60°C, 0 to 80% R.H.
(batteries not fitted)
Temperature oefficient:
0.15 x (Spec. Accy) / °C, < 18°C or > 28°C
Measure
Samples 3 times per second normal
Altitude
6561.7 ft. (2000m)
Safety
Complies with EN61010-1, UL61010-1,
IEC61010-1, CAT IV 600V, CAT III 1000V
Equipment protected by double insulation.
Instrument has been evaluated and complies with insulation
(overvoltage) catergory IV. Pollution degree 2 in accordance with
IEC-644. Indoor use.
CAT
Application FIeld
I
The circuits not connected to mains
II
The circuits directly connected to
Low-voltage installation
III
The building installation
IV
The source of the Low-voltage installation
Compliance to EN 61557 : IEC61557-1, IEC61557-2,
IEC61557-4, IEC61557-10