RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
flicker emissions
IEC 61000-3-3
GUIDANCE AND MANUFACTURE'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The BW-TSX / TS1 is intended for use in the electromagnetic environment specified below.
The customer or the user of BW-TSX / TS1 should assure that it is used in such an
environment.
IMMUNITY
IEC 60601 TEST
LEVEL
TEST
Electrostatic
±6 kV contact
discharge
±8 kV air
(ESD)
IEC
61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC
±1 kV for input/
61000-4-4
output lines
28
BewellConnect - BW-TSX/BW-TS1 - User Manual - 11-2017v2
The BW-TSX / TS1 use RF
energy only for its internal
function. Therefore, its
RF emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment.
The BW-TSX / TS1 is suitable
for use in all establishments,
including domestic
establishments and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
COMPLIANCE
ELECTROMAGNE
LEVEL
TIC ENVIRONMENT
- GUIDANCE
±8 kV contact
Floors should be
±15 kV air
wood, concrete or
ceramic tile. If floor
are covered with
synthetic material,
the relative humidity
should be at least
30%.
±2kV for power
Mains power quality
supply lines
should be that of a
typical commercial
or hospital
environment.
Surge
± 1 kV line(s) to
IEC
line(s)
61000-4-5
± 2 kV line(s) to
earth
Voltage dips,
<5% U
T
short
(>95% dip in U
T
interruptions
for 0.5 cycle
and voltage
40% U
T
variations on power
(60% dip in U
)
T
supply input lines
for 5 cycles
IEC 61000-4-11
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
T
for 5 sec
Power
3 A/m
frequency
(50Hz/60Hz)
magnetic field
IEC
61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
GUIDANCE AND MANUFACTURE'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The BW-TSX / TS1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BW-TSX / TS1 should assure that it is used in such an
environment.
IMMUNITY
IEC 60601 TEST
LEVEL
TEST
BewellConnect - BW-TSX/BW-TS1 - User Manual - 11-2017v2
±1 kV differential
Mains power quality
mode
should be that of a
typical commercial
or hospital
environment.
<5% U
Mains power quality
T
)
(>95% dip in U
)
should be that of a
T
for 0.5 cycle
typical commercial
40% U
or hospital
T
(60% dip in U
)
environment. If the
T
for 5 cycles
user of the BW-TSX
70% U
/ TS1 requires
T
(30% dip in U
)
continued operation
T
for 25 cycles
during power mains
<5% U
interruptions, it is
T
)
(>95% dip in U
)
recommended that
T
for 5 sec
the BW-TSX / TS1
be powered from
an uninterruptible
power supply or a
battery.
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
COMPLIANCE
ELECTROMAGNETIC
LEVEL
ENVIRONMENT - GUIDANCE
29