Cleaning and Maintenance
Maintenance
Inserting the new Hepa filter
1.
Place the new Hepa filter on top of the guiding rails with the label containing the serial number facing towards you. Make sure, the sides of
the Hepa filter and the guiding rails are aligned.
2.
Slide the new Hepa filter back into the housing.
3.
In the 48" version you replace two Hepa filters side by side.
4.
Make a note of the serial number at the front of the Hepa filter. You will need it later for the life time surveillance activation.
5.
Push back the lifting helps of the Hepa filter housing into their original position.
The housing now rests directly on top of the isolation foam of the filter.
6.
Secure the housing by retightening the two black screws on the left and right side between Hepa filter housing and walls of the top housing.
7.
Make sure that the Hepa filter is placed flush inside the housing and cannot be moved.
If secured tightly, the isolation foam on top and bottom of the filter will now prevent any airflow leakage.
8.
Close the housing front door. To do so, first unlock the two lifting helps on the right and the left side by slightly lifting the front door on
both sides upwards.
9.
Switch on the power.
The front door is equipped with a safety switch that automatically interrupts the power. If the touchscreen does not work after switching
on the power, ensure that the front door is fully closed.
Fig. 50: Activate the automatic life time surveillance
40
· GuardOne
· Manual
®
Activating the automatic life time surveillance
1.
To get to the Home page, press Tap to unlock.
2.
Press Settings.
3.
Press Replace Hepa Filter.
4.
Enter the 10-digit serial number (1) shown on the front of the
filter labels.
5.
Press Enter (2) to finish the installation procedure.
The GuardOne
Workstation is now ready to be used.
®
The remaining life time of the Hepa filter is 600 days.