CLEANING AND MAINTENANCE

AIR FILTER

Your Ombra* Table Top Compressor is equipped with a special air filter that will slowly change colour with use. The original
filter is bright white, but it will slowly turn grey with use. Your air filter should be changed every 6 months, or more often as
visual inspection indicates. The compressor should never be used without an air filter installed.
To change the filter (Figure 8)
1. Remove
AIR FILTER COVER
cap and filter cover out.
2. Remove filter from the cover, and dispose of the old filter.
3. Insert a new filter into the cover, then place the
approximately 90 degrees to tighten.
4. Do not over tighten. There will be resistance when the
FIGURE 8
90˚
CAUTION: Do not wash or reuse the filter. Use only Ombra* Replacement Filters (Part Number: 105544-001). Do not
attempt to substitute any other material to act as a filter as this will result in damage to the compressor and may void your
warranty.

NEBULIZER

OWNER'S MANUAL
Refer to the
clean your nebulizer.

EZ TWIST TUBING

Occasionally you may notice condensation on the inside of your oxygen tubing. This is normal. To remove the condensation,
simply connect one end of the EZ Twist
without the nebulizer attached. This will force air through the tubing, drying the inside. To clean the exterior, wipe with a damp
cloth. Ensure the tubing is dry prior to storage.

COMPRESSOR

Prior to cleaning the compressor, ensure the unit is unplugged from the electrical outlet. Use a damp cloth to wipe the
exterior of the compressor housing. Never use liquid or spray cleaners directly on the compressor.
CAUTION: Liquids can cause damage to the electrical components inside your compressor and lead to malfunction.
In the event any liquid enters the unit, please contact your Ombra* Authorized Distributor. Do not use the compressor until
it has been inspected, even if it appears to be functioning properly.

STORAGE

The Ombra* Table Top Compressor and reusable AEROECLIPSE* XL BAN should always be stored together with all of the
parts in a dry, dust-free environment.
DANGER: Do not attempt to disassemble the compressor housing as this presents an electric shock hazard. There are
no parts inside the compressor housing that require maintenance or replacement. Should you require repair or service for
your compressor, please contact your Ombra* Authorized Distributor.
WARNING: No modification of this equipment is allowed.
EN
8
by holding the cover and turning counterclockwise approximately 90 degrees. Pull the
AIR FILTER COVER
included with your reusable AEROECLIPSE* XL BAN for instructions on how to properly
TUBING
AIR OUTLET
to the
back into the compressor and turn clockwise
AIR FILTER COVER
is securely fastened.
on the compressor and allow the compressor to run
90˚