Instructions specific to hazardous area installations
(Reference European ATEX Directive 94/9/EC, Annex II,
1/0/6)
The following instructions apply to equipment covered by certificate number DMT 01
ATEX E 038:
1.
For use and assembly, refer to the main instructions.
2.
The equipment is certified for use as Category 1/2 D equipment. The Essential
Health and Safety Requirements are assured by compliance with EN 50281-1-1:1998;
Dust Explosion Protection.
3.
The equipment may be used with dust and fibers with apparatus temperature class
T. See table below.
4.
Thermal Data:
a. For 7ML5421 Series (Liquids version)
Permitted ambient temperature at the horn antenna (Category 1D):
- 20 °C
Permitted ambient temperature at the electronic enclosure (Category 2D):
-20°C
Maximum permitted temperature within electronic enclosure (Category 1D):
85 °C
Ambient temperature at the
horn antenna (Category 1D)
- 20 °C
T
amb
- 20 °C
T
amb
- 20 °C
T
amb
Page EN-8
SITRANS LR 400 – QUICK START MANUAL
T
+250 °C
amb
T
+65°C
amb
Maximum surface
temperature at the
horn antenna
(Category 1D)
72 °C
+ 60 °C
112 °C
+ 100 °C
262 °C
+ 250 °C
Maximum surface temperature Category 2D
(electronic enclosure resp. flange)
70 °C at T
65 °C in Category 2
amb
100 °C independent from T
250 °C independent from T
in Category 2
amb
in Category 2
amb
7ML19985QN83