HOW TO CHANGE THE BATTERY

1.
Hold the handpiece in the palm of your left hand.
2. Use your right hand to turn the bottom of the handpiece approximately 10 degrees anti-clockwise (twist then pull, Figure 9).
3. Carefully remove the battery from the handpiece.
4. Insert a new battery and lock in position twisting it 10 degrees clockwise (push then twist, Figure 10).
5. Turn ON the Radii Xpert handpiece and place it into the charging dock until status ring shows solid green.
Note: The LED attachment does not have to be attached to the handpiece to charge the battery.
6. Dispose of used battery in accordance to local regulations. Do not dispose with unsorted household waste.
Battery Life Expectancy: Battery life is dependent on the amount of use and LED efficiency. To extend the life of your Radii Xpert battery, keep it on
the charging dock at the end of each day if possible. In doing so SDI Ltd would expect consumers to experience a minimum of 300 total discharges.
TROUBLE SHOOTING
PROBLEM
Handpiece status ring flashes red.
Handpiece status ring shows solid red.
Handpiece status ring flashes blue once and
goes OFF.
No light appears on the charging dock when
handpiece is docked.
Unit works fine yesterday but will not turn on
this morning.
No audio beep heard when connecting with
the handpiece.
The light intensity is too low.
The LCD on the handpiece flashes.
Battery shows corrosion or emits a scorched
odour.
The dental material does not
cure completely.
I want to delete my favourite setting and go
back to the factory setting.
The handpiece has been on the charger
overnight. Status ring still showing red and
handpiece felt very hot.
The light emission does not start when the
button is pressed.
The LED attachment does not fit
the handpiece.
6
POSSIBLE CAUSE
Handpiece over heated.
Battery low.
Battery empty.
No power.
Moisture from condensation.
Battery empty.
1. Dirty or scratched lens cap.
2. Barrier sleeve.
Unit is in the setting mode.
Do not use unit.
Wrong attachment?
Charging problem. Battery maybe at the end
of its useful life.
Battery is flat.
Wrong Radii LED attachment.
SOLUTION
Allow handpiece to cool down before reuse.
Handpiece can still be used for curing. Return to
charging dock as soon as possible.
Return to charging dock for charging.
1.
Check all connections are clean and properly
connected.
2. Ensure power outlet is switched ON.
3. If the above does not work, move the whole unit
to a different powerpoint.
4. Call your local SDI Ltd representative for repair.
1.
Check for sign of moisture inside the barrier
sleeve.
2. Remove barrier sleeve and allow unit to dry for
a few hours.
Return to charging dock for charging.
Check and clean lens cap. Replace if required.
Ensure barrier sleeve fits tightly over the emission
tip.
Press
once to come out of the setting mode.
Call your local SDI representative.
Ensure the correct LED attachment is used. Check
material specifications for curing requirement from
the manufacturer.
Place the handpiece on the charging dock, wait for
status ring to show charging status
Press
(top & bottom button)
simultaneously for 1–2 seconds, then status ring
illuminates solid yellow, keep holding until you hear
one long beep which confirms Unit has returned to
factory setting.
Replace battery.
Return to charging dock for charging.
LED attachments from previous model of
Radii is not compatible with the Radii Xpert.