Télécharger Imprimer la page

First Use; Cleaning And Disinfection; Manual Cleaning - HEBU medical 1/1 Mode D'emploi

Publicité

Les langues disponibles

Les langues disponibles

Container systems
4. throw away the disposable filter(s)
5. remove disposable locks and indicators
!
All paper-filters are for single use only and should be thrown away
after each processing-cycle
Paper Filters
HE 3010-01
HE 3040-01
HE 3050-01

5.3 First use

The new container needs proper cleaning before first use.
The container has come to terms with a validated, automated cleaning and disinfecting
program.
In the machine to use there should be a neutral detergent (pH 7). Similarly, even a
mildly alkaline cleaner (pH 10-11) are used.
After processing in the cleaning and disinfection equipment, we recommend the
sterilization of the container in 121°C our 134°C
We also recommend the maintenance of all moving parts on the container with
instrument oil.
After cleaning, a new filter must be inserted (see 5.6

5.4 Cleaning and disinfection

Note: Improper cleaning and desinfection can lead to corrosion and stress cracks. Therefore
follow the specifications the cleaning and disinfecting agent's manufacturer recommends.
Detergents must be alkaline-, sodium- and carbonate-free, of neutral pH (7) and mild. Only
fully deionised water (quality according to EN 285 Annex B) is recommended for
reprocessing the product.
Before every use, the container has to be cleaned and disinfected.
Removal containers have to be cleaned and disinfected each time after use.

5.4.1 Manual cleaning

Mild detergents or neutral cleansers that are specially recommended for anodised
aluminum by the manufacturer are to be used in combination with a soft sponge and
water.
PTFE Permanent Filter
HE 3010-06
HE 3040-06
HE 3050-06
Filter change
EN
)
23

Publicité

loading

Ce manuel est également adapté pour:

3/41/2