11. Reactivation of temperature monitor

In order to avoid an overheating and
consequently the destruction of impor-
tant Polymax components, all units have
been equipped with a temperature con-
trol (11). This part is situated from out-
side at the bottom of the polymerisa-
tion pot and interrupts the circuit of the
heating in case of too much tempera-
ture development. In order to reacti-
vate the temperature monitor discon-
nect the unit from the mains and let it
cool down. Drain the water completely
through the drain tap (23). Lay the Poly-

12. Additional fields of application

All Polymax units are suitable for pres-
sure compression of embedding masses,
plasters and silicones. For this purpose
drain water through the drain tap (23),
close the tap again and adjust the tem-
perature to 0 °C (32 °F) (see 7.1).
Important: Heating up the poly-
merisation pot without water may
lead to damage of the heating
cartridge.
44
max on its side and use a soft tissue as
underlay, if necessary.
Unscrew the perforated metal plate at
the bottom of the unit and push the red
reset knob of the temperature monitor
until it latches completely. Insert per-
forated metal plate again and tighten
it. Put the Polymax into normal posi-
tion, close drain tap (23), fill the unit
with water and connect to the mains by
means of the power cord. Then go on
as described from point 7 on.
Open the lid (2) at the handle (1) and
put the objects into the dry polymeri-
sation pot.
Tip: In order to avoid impurities and/or
unevenness at the bottom of the pot
caused by plaster or embedding masses
we recommend an underlay of paper
or the like.
Close the lid of the pot up to the block
and start pressure-polymerisation
(see 7.3).

13. Care and maintenance

Disconnect the unit from mains sup-
Attention: Do not use scouring
ply before care and maintenance! The
agents, steel-wool or sponge
water inside the pressure pot has to
scourers.
be changed regularly (at least once a
week), because the lime particulates
Clean the interior pot with fresh water,
harm the surface of the pressure pot.
a neutral washing-up liquid and a soft
scarf or sponge. Afterwards let the
Polymax 1, 3 and 5
water run out. Close the drain tap (23)
For the cleaning of the pressure pot
and fill with fresh water. The O-ring
open the drain tap (23) at the rear of
should be greased regularly with sili-
the unit and let the water flow out.
cone grease (35 g tube,
Remove dirt thoroughly.
D3350).