Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
Polymax 1, 1 / 120 °C, 3 and 5
5. Initiation
Before initiation make sure that the volt-
age indicated on the type plate meets
the voltage of the power source.
When selecting a suitable place for
the machine, please make sure that
the storage space is even, resistant to
splash water as well as sufficiently sta-
ble for a minimum weight of 21 kg.
Connect the unit to an air compressor.
Push the pressure hose (included in the
delivery) into the compressed air con-
nection (25) until it is locked in place.
The hose does not need any further
fastening. To release the pressure hose
push the blue ring upwards and draw
the hose off by pulling downwards.
The pressure of the compressed
air line must not exceed 10 bars.
38
Put the power cord into the female con-
nector (15) and create a connection with
the Euro / USA plug. Open the lid (2)
by means of the handle (1) and fill the
unit with water. (Maximum water level:
30 mm below the upper edge).
Important: Before first initiation please
read the working instructions thor-
oughly. In case of comprehension prob-
lems contact your responsible dealer or
Dreve, if necessary.
Scope of delivery
1 Polymax
1 power cord
1 air connection hose incl. hose
clamps and fast coupling
1 water drain hose incl. coupling
1 working instructions in 6 languages

6. Keyboard unit

c
d
e
f
g
No.
Description
a
Manometer
b
Adjustment knob
c
LED – heating
d
Temperature indication display
e
Temperature switch
f
Time indication display
g
LED – polymerisation
h
only Polymax 1 / 120 °C – lid opening control
a
b
h