Instructions for Use
Installation
The turbopump with ISO F inlet flange must be fixed to the vacuum
chamber by means of bolts with a strength class of 12,9
(σr = 1200 N/mm
of steel bolts and the relevant fixing torque as recommended by
Agilent.
Tab. 1
FLANGE
ISO 250 F
WARNING!
The Turbo-V 2300 TwisTorr cannot be fixed by means of its base. The system
can only be fixed through its ISO 250F flange.
WARNING!
Failure to comply with these installation instructions could result in the pump
detaching from the system in the event of a rotor failure, which could cause
property damage or serious injury or death.
For installation of optional accessories, see "Technical Information".
234/262
2
). The following table shows the necessary number
FIXING DEVICE
M10 bolts
Turbo-V 2300 TwisTorr Pumps User Manual / 87-901-017-01 (C)
N.
FIXING TORQUE
12
30 Nm