1. Press MODE until mV values are shown. The display will always show the raw
signals of the electrodes in mV. Select the desired channel with the arrows.
2. Should you want to recalibrate, press CAL.
3. The display shows all the standards in memory:
pH: 1.68, 4.00, 4.01, 6.87, 7.00, 9.18, 9.21, 10.00, 12.45, Manual.
Ion: 1, 10, 100 ng/l, 1, 10, 100 µg/l, 1, 10, 100 mg/l, 1 g/l.
O
: 100% saturation.
2
mV: adjust potential.
°C: adjust temperature.
Select the proper value suitable for to the connected electrode and press
CAL.
4. pH:
rinse the electrodes with distilled water and immerse them in the first stand-
ard solution. Press CAL and wait until the measurement is stable (mV stops
blinking). Press CAL to confirm. Repeat eventually the same steps with a sec-
ond standard solution.
mV or °C:
adjust the value with the arrows and press CAL.
Ion:
rinse the electrodes with distilled water and immerse them in the first stand-
ard solution. Press CAL and wait until the measurement is stable (mV stops
blinking). Press CAL to confirm. Repeat the same steps with a second standard
solution. Eventually perform a blank correction after the second calibration.
O
:
2
hold the electrode in the air and wait until the measurement is stable (mV
stops blinking). Press CAL to confirm.
5. After rinsing the electrodes with distilled water, immerse them in the sam-
ples.
6. Rinse the electrodes always with distilled water after use and store them
according to the maintenance instructions.
1. Press MODE.
2. Select [INFO] to see general information about the instrument.
D130 èwww.consort.be