Specifications - DeVilbiss Pulmo-Aide 5650C Guide D'instructions

Masquer les pouces Voir aussi pour Pulmo-Aide 5650C:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12

SPECIFICATIONS

Size 3655 Series...................7.5" W x 4.0" H x 7.2" D (19.1 cm x 10.2 cm x 18.3 cm)
4650 Series ..............................7" W x 3.8" H x 13" D (18 cm x 10 cm x 33 cm)
5650 Series .......................10.1" W x 10.5" H x 6.5" D (26 cm x 27 cm x 17 cm)
Weight 3655 Series.....................................................................4.8 lbs. (2.18 kg)
4650 Series.....................................................................5.3 lbs. (2.40 kg)
5650 Series ......................................................................7.1 lbs. (3.2 kg)
Electrical
Requirements 3655D, 3655T, 4650D & 46502 ...........2.5 amps, 115 VAC (±10%) 60 Hz
3655I & 4650I.................................1.4 amps, 230 VAC (±10%) 50 Hz
3655K & 3655P................................1.4 amps, 230 VAC (±10%) 60 Hz
5650D, 5650H & 5650C ..................................1.3 amps, 115VAC 60Hz
5650I ..........................................................0.8 amps, 230VAC 50Hz
5650K & 5650P ........................................0.8 amps, 230VAC 50/60Hz
Sound Level 3655 Series ...........................................................................58 dBA
4650 Series............................................................................53 dBA
5650 Series............................................................................51 dBA
Power Consumption 3655D, 3655T, 4650D & 46502.....................140 watts maximum
Maximum Compressor Pressure 3655 & 4650 series .......35 psig or greater (241 kPa)*
Compressor Free Air Flow 3655 & 4650 ........................................8 lpm or greater*
Operating
Temperature Range all Models except 5650C ................+40° to +104°F (+5 to +40°C)
Operating Humidity all Models .......................................up to 95% non-condensing
Storage/Transport Temperature Range all Models ......-40° to +158°F (-40° to +70°C)
Storage/Transport Humidity all Models............................up to 95% non-condensing
Supplied Nebulizer Capacity all models....................................................6 ml (cc)
Supplied Nebulizer Nebulization Rate all models.......................0.15 ml/min or greater*
Supplied Nebulizer MMAD all models.........................................5 microns or less**
*Conditions may vary based on altitude above sea level, barometric pressure, and temperature.
**Value determined with minimum flow rate through the nebulizer (20°C, sea level).
There are no hazardous materials used in the construction of this unit, therefore, there
are no known risks associated with disposal of the equipment.
This product (models 3655D & 3655T) complies with the following electromagnetic
compatibility standard: IEC 60601-1-2:2nd Edition
3655D & 3655T Equipment Classifications:
Protection against electric shock:................................................................Class II
Degree of protection against electric shock: .................................................Type BF
Equipment not suitable for use in the presence of flammable anaesthetic mixture with
air or with oxygen or nitrous oxide
Mode of operation:.................................................................Continuous operation
A-655
3655I, 3655K, 3655P & 4650I .....................160 watts maximum
5650D, 5650H & 5650C ................................90 watts maximum
5650I, 5650K & 5650P ...............................100 watts maximum
5650 series ..................30 psig or greater (206 kPa)*
5650 ...................................................9 lpm or greater*
5650C ......................................+40° to +101°F (+5 to +38.5°C)
S p e c i f i c a t i o n s

Publicité

Table des Matières
loading

Table des Matières