Extraoral Suction System, Instructions For Use and Installation
UV-C Light Replacement
7.6
The UV-C light have a life time of 8,000 hours, please check them on a regular basis especially
when your are changing the filters that they are fully functional.
•Turn off the systems power and remove powercord from wall.
•Keep your hands dry.
•Wear gloves.
① Remove door panel.
② Unlock the dust-proof cabinet drawer and replace the cover plate.
③ Remove the filter box.
④ Turn over the filter box, making sure not to
damage the face of the HEPA filter.
⑤ Remove the UVC light tube and replace it with
a new one, dispose it as medical waste.
(UV light without mercury)
⑥ Remove the filter box and put the cover plate
back on. Double check and lock.
⑦ Reinstall the door panel.
[Caution]Replacement UV –C
Light should be disposed of as
medical waste.
[WARNING! ]
• •Do not look at UV lamps without
using proper eye protection.
• •At all times wear gloves when
handling UV lamps and quartz
sleeves, as oils from skin will
decrease UV energy transmission
once the lamp is heated and may
lead to premature failure.