Loosen the safety screw on the upper
side of the handle. Remove the KaWe
PICCOLIGHT F.O. – otoscope head
from the handle and pull out the lamp.
If necessary, clean the glass bulb of the
new lamp with alcohol. The bulb should be clean and free of finger
prints and oil or grease. The new lamp must be inserted all the way.
Loosen the safety screw on the upper
side of the handle. Unscrew the KaWe
PICCOLIGHT – ophthalmoscope
head from the handle. Remove the used
lamp and reinsert a new one, ensuring
that it is pushed in all the way.
Technical Specifications for the KaWe MedCenter 5000:
Power plug-AC adapter ............................Type UES12LCP-055200SPA
Wide range input voltage ................................100-240 V~, 50/60 Hz
Output voltage ......................................................................... 5.5 V DC
Output current ........................................................................ max. 2 A
Cord length ......................................................................................3 m
Protection class ....................................................................................II
Base module (with frame):
Dimensions of the base module ...................204.5 x 96.5 x 135.5 mm
Spiral power cord length .............. 2800 mm (700 mm coiled length)
Weight ....................................................................................ca. 810 g
Extension module:
Dimensions of the additional module...............80 x 96.5 x 135.5 mm
Spiral power cord length ................2800 mm (700 mm coiled length
Weight ....................................................................................ca. 370 g
Ear funnel module:
Dimensions of the ear funnel module...............80 x 96.5 x 135.5 mm
Weight ....................................................................................ca. 200 g
Environmental condition limits during use:
Temperature .......................................................... +10 °C to +40 °C
Relative humidity ........................................................... 10% to 75%
Air pressure ....................................................... 700 hPa to 1060 hPa
Storage and transport condition limits:
Temperature ........................................................... -10 °C to +60 °C
Relative humidity ............................................................10% to 95%
Air pressure ....................................................... 700 hPa to 1060 hPa
Warning: This device is maintenance free. For this reason, the hous-
ing can only be opened on the side end of the module's frame for the
addition of add-on parts. Repairs are only to be carried out by a qua-
lified expert. Never use more than two handles for each KaWe Med-
Center 5000 at the same time! The unit does not have any replacement
parts.
If damage is apparent, disconnect the power cord immediately. The
adapter serves in this case as the separator and must be disconnected
from the mains supply if damage is apparent. The adapter must be
easily accessible. The charging station is not designed for use in en-
vironments in which explosions may occur, in the presence of strong
magnetic fields, or in surgery. The spiral cable can cause people to trip.
Do not place it where people may be walking. Only KaWe-brand com-
ponents are to be used. Modifications to this device are not allowed.
While performing examinations, the user is not to touch the
conducting parts such as the electric plug, screw closure on the
diagnostic unit, etc.
When the light is left on for longer than 1 minute, the metal on the
upper part of the handle becomes very hot and when touched can
hurt or cause burns. If the handle left on surfaces that are temperature
sensitive, there may be a risk of fire. Ensure that after each use, the
handle is returned to its holder and the light is turned off. After using
for an extended period of time, allow enough time for the handle to
cool.
In order to prevent heat buildup on the top part of the handle, the unit
turns itself off automatically after 3 minutes. Please activate by press-
ing the button on the handle or returning to the wall charging station.
otoscopes/
KaWe
ophthalmoscopes are to be used at room temperatures
otoscopes/
of not greater than 40° C. KaWe
to special safety measures with regard to their electromagnetic compati-
bility. The operation of the KaWe
fluenced by portable and mobile high-frequency communication devices.
Instructions
Storage and Transport: Please store this device in its packaging
prior to assembly in order to avoid contact with impurities or dust.
Cleaning Preparation: No specific requirements.
Manual Cleaning: The external parts of the charging station and
the diagnostic equipment can be cleaned with a damp, soft, lint-free
cloth. Soapy water or other non-aggressive agents may be used.
Automatic Cleaning: No automated cleaning possible / necessary.