DO NOT use this product in or near explosive environments, or where aerosol
(spray) products are being used.
12.
DO NOT pump combustible liquids or vapors with the product or use in or near an
area where flammable or explosive liquids or vapors may exist.
13.
DO NOT operate this product in an oxygen enriched environment, i.e. oxygen tent
or oxygen hood.
14.
Use only in well ventilated areas.
Safety Overload
NOTE
Do not exceed 30 minutes of continuous operation.
NOTE
This is based on the thermal interrupt safety device that is integrated into the mo-
tor. As the unit runs it generates heat and gets warm. The thermal interrupt will cut
power to the unit if the temperature inside gets to an unsafe level. Once the unit
cools off it will work normally. If the unit is run for over 30 minutes continuously the
thermal interrupt may trip which would stop the user's treatment
1.
The motor of this device has a built-in thermal overload protector. Should the mo-
tor overheat, the protector will automatically shut off the motor. Should this occur,
turn the I/O (ON/OFF) switch to the O (OFF) position and allow the motor to cool
down for approximately 30 minutes before turn it on again.
2.
If the overload protector shuts off the motor frequently, you may have an unstable
voltage situation.
Rev.4.07.28.15
10
............................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL MQ5800, MQ5900
............................................................................................................