Télécharger Imprimer la page

Beurer medical BM 55 Mode D'emploi page 29

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 31
• the pump pressure is higher than 300 mmHg (
on the display)
• pumping up takes longer than 160 seconds (
the display)
• there is a system or device error (
pears on the display)
• the batteries are almost empty
• the data could not be sent to the PC (
display).
In such cases, repeat the measurement. Ensure you do not
move or speak. If necessary, re-insert or replace the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical specifica-
tions", the technical alarm will appear on the display indicating
either " " or " Lo Lo ". In such cases, you should seek medical as-
sistance and check the accuracy of your procedure.
The limit values for the technical alarm are factory set and can-
not be adjusted or deactivated. These alarm limit values are
accorded second priority under the standard IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear automati-
cally after about 8 seconds.
11. Specifications
Model no.
BM 55
Type
M1002
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the upper arm
appears
appears on
,
,
or
ap-
,
appears in the
Measurement range
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operating
conditions
Permissible storage
and transport condi-
tions
Power supply
Battery life
Accessories
Classification
29
Cuff pressure 0 – 300 mmHg,
systolic 60– 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 186 mm x W 95 mm x H 56 mm
Approx. 467 g (without batteries)
22 to 36 cm
+10 °C to + 40 °C, ≤ 90 % relative air
humidity (non-condensing)
- 20 °C to + 55 °C, ≤ 90 % relative air hu-
midity, 800 –1050 hPa ambient pressure
4 x 1.5 V
AAA batteries
For approx. 200 measurements, de-
pending on the blood pressure level
and/or pump pressure
Cuff, instructions for use, 4 x 1.5 V AAA
batteries, USB cable, storage pouch
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF

Publicité

loading