CONTRAINDICATION

It is inappropriate for people with serious arrhythmia to use this Electronic Sphygmomanometer.

PRODUCT DESCRIPTION

Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure and pulse
rate can be measured automatically and non-invasively. The LCD display will show blood pressure and
pulse rate. The most recent 2x60 measurements can be stored in the memory with date and time stamp.
The monitor can also show the average reading of the last three measurements.
The Electronic Sphygmomanometers corresponds to the below standards: IEC 60601-1:2005/EN 60601-
1:2006/AC:2010 (Medical electrical equipment -- Part 1: General requirements for basic safety and essential
performance), IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical electrical equipment -- Part 1-2:
General requirements for basic safety and essential performance - Collateral standard: Electromagnetic
compatibility - Requirements and tests), EN 1060-1: 1995 + A1: 2002 + A2: 2009 (Non-invasive
sphygmomanometers - Part 1: General requirements), EN 1060-3: 1997 + A1: 2005 + A2: 2009 (Non-invasive
sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood pressure mea-
suring systems). ANSI/AAMI SP-10: 2002+A1:2003+A: 2006.