Dreve Durfosmart scan Mode D'emploi page 19

Thermoformeuse universelle sous pression 4 bars
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 24
Drufosmart scan
9. Deviation to the rule
9.1 Modification of given times
Scanned times may be prolonged or
shortened individually, if required.
Info
The performance of the heating
might decrease slightly after 1000
operating hours. By prolonging the
heating times by 5–10 seconds
optimum thermoforming results
can be attained again. The times
given by the barcodes are average
times and designed for general
use. In case of extreme undercut
areas or very high models a pro-
longation of the heating time can
optimise the results.
After having scanned the pre-adjusted
material parameters (see paragraph
8.1) the heating time (1 x) or the cool-
ing time (2 x) can be selected by press-
ing the menu key (j). There will be an
arrow in the display in front of the cho-
sen time.
The heating time can be modified posi-
tively and negatively in steps of 5 sec-
onds. For this press the menu keys (e/i)
according to the key description. The
max. deviation from the scanned time
is 35 seconds.
36
With the same working steps the cool-
ing time can be modified in steps of
10 seconds by max. 70 seconds.
For identification of the modified
parameters the display will show a +
resp. a - sign. The degree of deviation
will be shown by a scale at the bottom
of the display.
Modified values have to be confirmed
by pressing the OK key (f).
All further working steps correspond
to 8.2.
9.2 Creation and storage of manual time programmes
Up to three individual time programmes
For single use confirm the two entered
may be created and stored for thermo-
values by pushing OK (f). Then go on
forming foils that are not equipped with
as described in point 8.2.
a corresponding barcode.
If the created time programme is to
Turn on the unit – the initial display
be stored for repeated use, 3 memory
will be visible. When pushing the menu
capacities (F1–F3) are available.
key (j) one time the display will show
– – – – – at position b.
Do not confirm the created time pro-
gramme by pushing OK, but press menu
key (j) again. The display will show F1
in position b, the entered times will be
taken over. The memory capacities F2
and F3 can be selected by pushing the
menu key (j) 2 resp. 3 times. Store the
times in the desired memory capacity
As described in point 9.1, push menu
by pushing the OK key (f).
key (j) and select the desired heating
time by the menu keys +/- (e/i). Fol-
lowing push menu key (j) again and
enter cooling time.
Then go on as described in point 8.2.
37

Publicité

Table des Matières
loading

Table des Matières