SAFETY INSTRUCTIONS

4.1

General

The instrument HUMOR 20 is built according to the current state of the art and it
will operate reliably if it is free from faults and if it is properly operated and
maintained.
The instrument can present a risk if it is used improperly or not according to
specifications by unqualified personnel!
• damage to the instrument itself, to other assets of the
This can result in:
• inefficient and imprecise operation of the instrument.
For the sake of your own safety, the following tips in particular are to be
observed:
• Only qualified or specially trained personnel should be tasked with operating or
working with the humidity calibrator. Unauthorized changes or modifications to
the instrument are not permitted or require the express authorization of the
manufacturer.
• Carefully read through the operating instructions before starting up the instru-
ment. Furthermore, the supplied operating instructions must always be accessi-
ble to all personnel concerned with the transport, setup, start-up, operation, and
maintenance. (Be very careful when lending or disposing of the instrument to
third parties)
• The instrument should only be operated in a fault-free state. Possible defects
must be repaired either by authorized persons or by the customer service
department of E+E Elektronik ® Ges.m.b.H before starting up the instrument.
4.2

HUMOR 20

• Check whether the details for the supply voltage on the nameplate agree with
the local mains supply.
• Only use a grounded plug to connect the humidity calibrator to power
(protective measures).
• Before opening the water inlet, guarantee that the humidity calibrator is
no longer under pressure. (Controller turned all the way to the left, if HUMOR
is already in operation, display RH > 90% )
• The HUMOR should only be operated with distilled (deionized) water.
• Before connecting the compressed air, guarantee that the controller is turned all
the way to the left.
• The instrument should only be operated with filtered compressed air maximum
contamination level < 0.03 g/m³.
• The media supply should not exceed an absolute pressure of max. 10 bar.
(Manometer and safety valve in the supply line)
• A humidity calibrator that is ready for operation and that is filled with water
must not be tilted more than 20 degrees !
Otherwise empty HUMOR 20 totally --> page 27 : chapter 8.2
• Disconnect compressed air after power off (running risk of condensation in
the pipes because of switched off heating).
user, as well as to the operating personnel.