Instructions For Use - Sony HMM-3000MT Instructions D'utilisation

Masquer les pouces Voir aussi pour HMM-3000MT:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20
Precautions
Preparing for unforeseen circumstances
Be prepared for unforeseen circumstances by using a
stationary monitor, for example, in conjunction with this
unit. In addition, prepare a backup unit if necessary.
Use with electrosurgical knives and similar
devices
If this unit is used together with an electrosurgical knife,
etc., the picture may be disturbed, warped or otherwise
abnormal as a result of strong radio waves or voltages from
the device. This is not a malfunction.
When you use this unit simultaneously with a device from
which strong radio waves or voltages are emitted, confirm
the effect of this before using such devices, and install this
unit in a way that minimizes the effect of radio wave
interference.
Condensation
If the unit is suddenly taken from a cold to a warm location,
or if ambient temperature suddenly rises, moisture may
form on the outer surface of the unit and/or inside of the
unit. This is known as condensation. If condensation
occurs, turn off the unit and wait until the condensation
clears before operating the unit. Operating the unit while
condensation is present may damage the unit.
About the OLED panels
Due to the characteristics of the material used in the OLED
panel, permanent burn-in or reduction in brightness may
occur.
These problems are not a malfunction.
The OLED panel fitted to this unit is manufactured with
high precision technology, giving a functioning pixel ratio
of at least 99.99%. Thus a very small proportion of pixels
may be "stuck", either always off (black), always on (red,
green, or blue), or flashing. In addition, over a long period
of use, because of the physical characteristics of the
organic light-emitting diode, such "stuck" pixels may
appear spontaneously. These problems are not a
malfunction.
Be sure to read this document together with the
Instructions for Use for the HMI-3000MT.
18
Precautions / Specifications
Specifications
General
Power requirement
supplied from HMI-3000MT
Operating conditions
Temperature: 5 °C to 35 °C (41 °F to
95 °F)
Humidity: 25% to 80% (no condensation
allowed)
Pressure: 700 hPa to 1060 hPa
Storage and transport conditions
Temperature: –20 °C to +60 °C (–4 °F to
+140 °F)
Humidity: 20% to 80% (no condensation
allowed)
Pressure: 700 hPa to 1060 hPa
Dimensions (width / height / depth, including maximum
protrusions)
Approx. 191 mm × 173 mm × 271 mm
(7
in. × 6
5
8
Mass
Approx. 480 g (1 lb 0.93 oz.) (excluding
connection cable)
Display resolution (horizontal / vertical)
1280 × 720
Interpupillary distance (distance between eye pupils) range
53 mm to 76.7 mm (2
Supplied accessories

Instructions for Use (1)

Service Contact List (1)
Information for Customers in Europe (1)
Optional accessories
HMO-CA50M head mount display cable
Specifications and external appearance are subject to
change without notice for product improvement.
Medical Specifications
Protection against harmful ingress of water:
Ordinary
Degree of safety in the presence of a
flammable anesthetic mixture with air or
with oxygen or nitrous oxide:
Not suitable for use in the presence of a
flammable anesthetic mixture with air or
with oxygen or nitrous oxide
Mode of operation:
Continuous
in. × 10
in.)
7
3
8
4
in. to 3
in.)
1
1
8
8

Publicité

Table des Matières
loading

Table des Matières