Télécharger Imprimer la page

Biomet Certain ISRT10 Mode D'emploi page 3

Kit de retrait de vis cassées

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 8
Step 4: Remove the ISRT07 Manual
ISRT07 Manual
Reversing Drill
ISRT05 Certain
®
Guide
Handle
IUNIHG Screw (Broken)
Step 5: Insert the ISRT06 Screw Removal
ISRT06 Screw
Removal Tool
OSSEOTITE
®
Certain
®
Implant
IUNIHG Screw (Broken)
Step 6: Remove the ISRT06 Screw
ISRT06 Screw
Removal Tool
IUNIHG Screw (Broken)
OSSEOTITE
Certain
®
®
Implant
en
Reversing Drill and then remove
the ISRT05 Guide Handle from the
implant.
Tool into the implant and seat the
tool onto the top portion of the
broken screw. Press onto the screw
with slight to moderate pressure to
engage/capture the screw. Rotate
the tool in the reverse direction
(counter-clockwise) several
rotations. The screw should release
from the implant internal threads.
Removal Tool from the implant.
The fractured screw should remain
captured in the ISRT06. The screw
can be removed by pulling it out of
the tool tip. If the tool fails to catch
the screw, repeat steps 2 – 5 until
the screw is extracted from the
implant.
IWSU30 Certain
®
Waxing Screw/Guide Pin
OSSEOTITE
®
Certain
®
Implant
ISRT01 Certain
®
Hand Tap
OSSEOTITE
®
Certain
®
Implant
Batch code
Caution, consult accompanying
documents
Consult Instructions for Use
Step 7: After the broken screw has been
removed, insert the IWSU30 Certain
Waxing Screw/Guide Pin to evaluate
the integrity of the internal threads
of the implant. If the IWSU30 does
not rotate easily into the implant,
proceed to Step 8. If the IWSU30
rotates easily into the implant, you
may proceed in placement of the
new abutment.
Step 8: If the IWSU30 did not rotate
easily into the internal threads
of the implant, insert the ISRT01
Certain Hand Tap into the implant.
Rotate the ISRT01 in a clockwise
direction to rethread the implant.
Before replacing the abutment,
checking the internal threads with
the IWSU30 Certain Waxing Screw/
Guide Pin is recommended. Once
the integrity of the implant threads
has been verified, the abutment
may be replaced.
Catalogue number
By prescription only

Publicité

loading