WARNING :
Heat development (burn hazard)
As it is the case with all high-performance lights, the high light intensity results
in a certain heat development. Pro-longed exposure of areas near the pulp and
soft tissues may result in irreversible or reversible damage.
Therefore, this high-performance curing light must only be operated by trained
professionals.
CAUTION :
When the curing light being used for over 10 seconds, it is recommended to
cool down for 5 minutes before starting the next curing cycle to avoid the
patient being burned by the high temperature fiber-optic probe.
(D) Using the Built-in Radiometer
1. Ensure the charger stand is plugged in.
2. Press the fiber-optic probe flat onto the radiometer sensor.
3. Start a curing cycle and take the light intensity reading.
Note: Any movement of the device while measuring the light intensity will result in
variations of the reading.
Note: To Change the temperature reading measurement unit, press the °C/°F switch.
(E) Handpiece Display Panel Descriptions
: Full Power Mode,the most using in clinical practice - to boost with the
full power light output for ultrafast performance to cure all resin-based
restoratives in 1, 2 or 3 seconds.
: Ramp-Up Mode, for lab use with automatic ramping up illumination to
minimizes composite shrinkage on curing the indirect restorations e.g.
crown, bridges, veneer, onlay or inlay.
: Pulse Mode,switching with high and low light output in less heat effect for
clinical flexibility using for curing liner, bonding adhesives and the resin
cement pre-cure.
: Blue light and UV light
UV
: Full Battery
: Battery Charging
: Overheating - Device will automatically shut off and will be able to
restart after cooling down.