Knick Stratos Pro A2 CONDI Série Guide Rapide page 10

Masquer les pouces Voir aussi pour Stratos Pro A2 CONDI Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 25
Calibration with Calibration Solution
Input of temperature-corrected value of calibration solution with
simultaneous display of cell factor.
This calibration is performed using known calibration solutions and
the respective temperature-corrected conductivity values (see table
on calibration solution). Make sure that the temperature does not
change during the calibration procedure.
Note:
When using an ARF 210/215 fl ow-through fi tting, you should use the
inluded calibration beakers (identical dimensions and materials) for
calibration to prevent calibration errors.
Display
10
Action
Select Calibration.
Press enter to proceed.
Select CAL_SOL
calibration method.
Press enter to proceed.
Ready for calibration.
Hourglass blinks.
Immerse sensor in
calibration solution.
Enter the temperature-
corrected value of the
calibration solution
using the arrow keys
(see table).
Press enter to confi rm.
Remark
Display (3 sec)
Now the device is in
HOLD mode.
Lower line:
display of cell factor
and temperature