Dentsply Sirona Purevac HVE Mode D'emploi page 6

Kit de démarrage avec adaptateur pour tuyau et canules d'aspiration de haut débit avec miroir
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 8
INSTRUCTIONS FOR REPROCESSING THE HVE MIRROR TIP
Reprocessing Step
Method
Sterilization
• Place each HVE Mirror Tip in a separate steam-sterilization pouch.
• Place bagged HVE Mirror Tip into a steam sterilizer, per the sterilizer
manufacturer's instructions.
• The following minimum sterilization cycles may be used:
• Gravity steam sterilization – full cycle: 135 °C for 10 minutes.
• Pre-vacuum steam sterilization – full cycle: 121 °C for 15 minutes.
• Pre-vacuum steam sterilization – full cycle: 135 °C for 3 minutes.
• Steam flush/pressure pulse sterilization (e.g. STATIM) – full cycle:
134 °C for 3.5 minutes.
• Alternate method: place non-bagged instruments into the steam
sterilizer and run at the above listed cycles.
11
Warning
Reprocessing Step
Before patient use, HVE
Point of Use
Mirror Tips must be cleaned
and sterilized.
Dentsply Sirona recommends
the use of Assure Plus
Self-
®
Cleaning &
Sealing Sterilization Pouches.
Disinfection: Manual
Note: Instruments sterilized
non-bagged should be used
immediately.
INSTRUCTIONS FOR REPROCESSING THE HVE HOSE ADAPTER
Method
• Turn suction off by switching the lever from on to off on existing HVE
valve.
• Remove the HVE Mirror Tip from the HVE Hose Adapter.
• To remove visible soil, thoroughly wipe the entire outer surface of
the HVE Hose Adapter, including the entire length of the hose and
the holster, with a new, single-use wipe in combination with an
alcohol-based, tuberculocidal, cleaning and disinfecting solution. Use
additional wipes as needed to ensure all visible soil is removed.
• Pay special attention to all seams and crevices while wiping.
• If soil is observed in the seam area of the HVE Hose Adapter,
disassemble the unit by unscrewing the top plastic component
(Figure 8), carefully removing the metal ring (Figure 9) and wiping
the components separately (Figures 10 and 11). Once visually clean,
place the metal ring back into the top plastic component and screw
the plastic pieces back together.
Figure 8
Figure 9
Figure 10
Figure 11
Warning
• Do not attempt to remove
the HVE Mirror Tip from the
HVE Hose Adapter while the
HVE system is operating.
• Use only a disinfecting
solution which is approved
for its efficacy, EPA
registered (and/or Health
Canada approved), and use
in accordance with the DFU
of the disinfecting solution
manufacturer.
Dentsply Sirona recommends
the use of VoloWipes
®
Disinfecting/Deodorizing/
Cleaning Wipes.
12

Publicité

Table des Matières
loading

Table des Matières