DIRECTIONS FOR USE

Cavitron® Ultrasonic Inserts are to be used with magnetostrictive
ultrasonic scalers.
Indications for Use
Used for ultrasonic procedures:
• All general supra and subgingival scaling applications.
• Periodontal debridement for all types of periodontal diseases.

Contraindications

Cavitron Inserts should not be used for restorative dental
procedures involving condensation of amalgam.
Warnings
• Tips with high usage which exceed the warranty period should be
2
discarded to avoid possible in-use breakage and possible injury
• Handle insert with care. Improper handling of insert, specifically
the insert tip, may result in injury and/or cross contamination.
• Failure to follow properly validated sterilization processes and
approved aseptic techniques for Cavitron Ultrasonic Inserts may
result in cross contamination.
• Cavitron® Ultrasonic Inserts with metal grips are not compatible
with the Steri-Mate® 360 Handpiece.
Precautions
It is the responsibility of the Dental Healthcare Professional to:
• Operate the inserts only at the recommended power settings.
• Check water spray to verify fluid is reaching the working end of the
insert tip. Adjust the water lavage control until the water (lavage)
flows with a rapid drip or small spray (approximately 35mL).
Higher water flow settings provide cooler irrigation.
• Not to bend or reshape insert. Inserts should be handled with care