Error Codes

Error Code
Error 1
Motor is not turning
Motor has failed to reach
the desired speed, or is
Error 2
drawing too much power
to reach speed desired.
Troubleshooting
Definition
Cause
Motor has overheated
Something is physically
preventing the motor
from turning
Generator probe is too large
for the speed desired or the
sample is too large/viscous to
be processed.
Possible Solution
Unplug and let stand
for at least 1 hour at
room temperature, then
attempt to restart.
Remove generator
probe and attemot to
run the motor.
Check that the
generator probe
(removed from the
motor) spins freely
Check for foreign
objects inside the
generator probe port.
Reduce desired speed
Reduce the generator
probe size
Reduce the viscosity
of the sample being
processed