Tableau 5-5 Messages d'erreur de l'auto-échantillonneur (suite)
Message d'erreur
Horizontal: home sensor activated
when not expected.
Horizontal: home sensor not
de-activated.
Horizontal: home sensor not
reached.
Horizontal: needle position is
unknown.
Illegal sensor readout.
Incorrect amount of steps executed
to reach the home position
Incorrect first destination vial.
Injection needle unit error.
Injection valve or ISS unit error.
Injection volume (number) is invalid.
For specified injection method,
volume should be within the range
%.2f µL-%.2f µL, with (number). µL
increments.
Injection volume error.
Invalid (number) vial position
(number). The vial position must be
between 01 and (number).
Guide de l'utilisateur du matériel
RUO-IDV-05-10983-FR-D
Description
• Mettez le module hors tension, puis sous tension.
Si le message s'affiche à nouveau, contactez
sciex.com/request-support.
• Vérifiez que le bloc d'aiguille n'est pas obstrué.
Mettez le module hors tension, puis sous tension.
Si le message s'affiche à nouveau, contactez
sciex.com/request-support.
• Vérifiez que le bloc d'aiguille n'est pas obstrué.
Mettez le module hors tension, puis sous tension.
Si le message s'affiche à nouveau, contactez
sciex.com/request-support.
• Initialisez le bloc d'aiguille à l'aide du logiciel.
• Mettez le module hors tension, puis sous tension.
Si le message s'affiche à nouveau, contactez
sciex.com/request-support.
• Vérifiez l'absence d'obstruction au mouvement
horizontal ou si le mouvement présente un couple
trop élevé.
• Corrigez les paramètres dans le logiciel.
• Vérifiez que le bloc d'aiguille n'est pas obstrué. Si
le message s'affiche à nouveau, contactez
sciex.com/request-support.
• Mettez le module hors tension, puis sous tension.
Si le message s'affiche à nouveau, contactez
sciex.com/request-support.
• Corrigez les paramètres dans le logiciel.
• Corrigez les paramètres dans le logiciel.
• Corrigez les paramètres dans le logiciel.
Dépannage
Système ExionLC 2.0