6.4 Series KERN ALS-A

 In weighing mode press MENU button. The first menu item „count" is displayed.
 Press MENU button
 Acknowledge using PRINT button, the current setting is displayed.
 Using MENU button select „d Liquid"
 Confirm using the PRINT button, the currently set density of the sinker is
displayed (factory setting 3.0000 g/cm
 For any change, enter the density of the sinker as follows.
To delete keep pressed the TARE button. Use the navigation buttons  to
increase/reduce the digit. Use the TARE button to select the next digit. Repeat
this sequence for each digit.
 Confirm input by pressing the PRINT button. The display for weight determination
of the "Sinker in air" appears.
 Confirm by pressing the PRINT button.
If weighing balance does not show Zero, press TARE.
 Place sinker into the upper sample dish, see fig. 5, chap. 6.1.2.
 Wait for stability display [ ], take over the weight value „sinker in air" using the
PRINT button.
 Wait until the display for weight determination of „sinker in test liquid" appears.
 Confirm by pressing the PRINT button.
 Place sinker into the lower sifting bowl, see fig.6, chap. 6.1.2..
For this remove the immersion basket out from the frame. Always ensure that,