ALARM TABLE

Display/Defect
AL 1 – Alarm heating element
broken
AL 2 – Alarm Low Temperature
AL 3 – Alarm Overheating
AL 4 - Probe broken
AL 5 – Black Out
Cleaning
ATTENTION: Before cleaning or screwing out any part of the machine make sure that this one is
disconnected from the feeding network
To make a good cleaning use only neutral detergents and rinse
The gaskets and all plastic parts can be cleaned with a cotton flock soaked in alcohol.
Do not use sponges or abrasive materials which can damage the surface.
To clean do not use detergents, disinfectants or other liquids containing chlorides as they are corrosive for
stainless steel; in case of contact please rinse immediately.
Warranty
This instrument is designed and built according to the Community Directive CEE 93/42 "Medical equipments"
and C.B.M. is responsible for safety, reliability and performances provided that possible modifications or
repairs are carried out by authorized people, the corresponding electric plant complies with IEC regulations
and the equipment is used in compliance with these instruction for use.
C.B.M. guarantees its incubator for one (1) year, in case of manufacturing defects or failures of components,
though not due, according to C.B.M. evaluation, to the operators' wrongful use of the apparatus.
The manufacturing company is released by any responsibilities for damages to people or things due to failure
to earth the plant.
The warranty does not cover parts subject to wear and repairs or replacements carried out by personnel not
authorised by C.B.M.
Details, descriptions and pictures are merely indicative.
C.B.M. S.r.l. reserves the right to carry out ameliorative modifications to its own products without any
previous notice.
Disposal Of Electric And Electronic Refuse
Waste from Electrical and Electronic Equipment - Directive 2002/96/EC (WEEE)
The symbol (shown above) that appears on this product means that it should not be disposed of together with
normal household refuse but should be treated separately.
Electric and electronic equipment may contain substances that are harmful to the environment and to the
health of humans and should therefore be disposed of at special waste disposal sites.
There are harsh legal penalties in force against persons found disposing of such equipment in an
irresponsible manner.
When the equipment is no longer of any use, it may taken back to the outlet from which you intend to
purchase a replacement.
Rev.3 – 28/07/2010
Cause
1) Excessive load
2) Heating element broken
Heating element broken
CPU card in Error
CPU card in Error
The Black Out
2429-2430-2431-V COMP- r3x4.DOC
Remedy
1) Reduce the load and repeat the cycle
2) Replace the faulty heating element
Replace the faulty heating element
1) Open immediately the door
2) Replace the faulty CPU card
Replace the faulty CPU card
Repeat the cycle